The new flashcard feature is useful to memorize questions.
?Change your study mode
At any time, you can change the study mode, and alternate between the practice mode and the exam mode. In practice mode, you can configure for example the number of questions or tests, and other parameters to help you study.
Randomized | 10 Questions per Test | 20 Minutes | 70% to pass|
To re-configure your study mode again and change - for example - the number of tests, whether you have random questions and all other configuration parameters.
?Simulator Configuration
Auto-scroll: You can use the automatic scrolling of the questionnaire that occurs as soon as you answer one or all of the answers to a question correctly. Auto scrolling is activated if you answer a single answer, or as soon as you answer all the mandatory answers. Learning Mode: During learning mode you can get a real time result for your answer.
Free Test
Question: / 10
20:00Min. left
?Restart the current test
To restart the current test by clearing all your answers and the time used up to now. Warning: all answers will be lost.
Question: / 10
5.0(67 Votes)
Quiz
Question 1/101/10
Cardiac Anatomy and Physiology
Cardiac Anatomy and Physiology
Cardiac Anatomy and Physiology
During implantation of a dual-chamber pacemaker, the atrial lead is to be positioned in the right atrial appendage. Where is the right atrial appendage located relative to the other right atrial structures?
Select the answer:Select the answer
1 correct answer
A.
Posteroinferiorly, adjacent to the coronary sinus ostium
B.
Anterosuperiorly, overlapping the root of the aorta
C.
Inferiorly, at the junction with the inferior vena cava
D.
Medially, within the fossa ovalis
E.
Posteriorly, between the pulmonary veins
Explanation: The right atrial appendage is a broad, triangular, trabeculated pouch that projects anterosuperiorly from the right atrium and overlaps the root of the ascending aorta. Its interior contains pectinate muscles, which create a lattice into which a passive-fixation J-shaped atrial lead can lodge securely. On fluoroscopy in the anteroposterior view, a lead tip in the appendage typically points medially and anteriorly with a characteristic to-and-fro "windscreen wiper" movement with atrial contraction. The coronary sinus ostium lies posteroinferiorly near the septal leaflet of the tricuspid valve, and the inferior vena cava enters the posterior inferior aspect of the chamber, guarded by the Eustachian ridge. The fossa ovalis is a depression on the interatrial septum and is a site used for transseptal access rather than lead fixation. The pulmonary veins drain into the left atrium, not the right. Understanding the anterosuperior position of the appendage helps the implanter interpret fluoroscopic projections, avoid placing the lead too laterally where the free wall is thin, and achieve stable sensing and capture thresholds with low dislodgement risk.
Right Answer: B
Quiz
Question 2/102/10
ECG and Arrhythmia Recognition
ECG and Arrhythmia Recognition
ECG and Arrhythmia Recognition
An ECG shows progressive PR prolongation followed by a non-conducted P wave, after which the cycle repeats. The QRS is narrow. What is the most likely level of block?
Select the answer:Select the answer
1 correct answer
A.
The His bundle below the penetrating portion
B.
The left bundle branch
C.
The right bundle branch
D.
The atrioventricular node
E.
The sinoatrial junction
Explanation: Progressive lengthening of the PR interval before a dropped beat defines Mobitz type I second-degree atrioventricular block, also called Wenckebach. With a narrow QRS, this pattern almost always reflects block within the atrioventricular node, whose decremental conduction properties produce incremental delay with each successive impulse until one fails to conduct. Other classic features include progressive shortening of the RR intervals during the cycle and a pause shorter than two PP intervals. Nodal Wenckebach is often vagally mediated, may occur during sleep or in athletes, and tends to improve with exercise or atropine. It generally has a benign prognosis when asymptomatic, and pacing is reserved for symptomatic patients or when block is shown to be infranodal. Intra-Hisian or infra-Hisian block more typically produces Mobitz type II, with constant PR intervals before sudden non-conduction, and is often associated with a wide QRS due to concurrent bundle branch disease. Sinoatrial exit block causes absent P waves rather than non-conducted P waves. Recognising the site of block guides urgency and the need for permanent pacing.
Right Answer: D
Quiz
Question 3/103/10
Electrical Principles and Device Technology
Electrical Principles and Device Technology
Electrical Principles and Device Technology
According to Ohm's law, what happens to current drain from a pacemaker output if lead impedance falls while voltage output remains constant?
Select the answer:Select the answer
1 correct answer
A.
Current decreases
B.
Current remains unchanged
C.
Current falls to zero
D.
Current increases
E.
Current becomes alternating
Explanation: Ohm's law states that voltage equals current multiplied by resistance, or V = I x R. Rearranging gives I = V divided by R. In pacing, impedance is the total opposition to current flow, encompassing conductor resistance, electrode-tissue interface resistance and polarisation. If the programmed voltage is constant and impedance decreases, more current flows with each pulse. This has two clinical consequences. Firstly, battery drain increases, shortening device longevity. Secondly, an abrupt fall in impedance, typically below about 200 ohms, suggests an insulation breach, which provides an alternative low-resistance path for current to leak, frequently between conductors in a bipolar lead or into surrounding tissue. Conversely, a sudden rise in impedance, often above 2000 ohms, suggests conductor fracture or a loose set screw, where current flow is reduced or interrupted, resulting in loss of capture. High-impedance leads were developed specifically to reduce current drain and extend battery life. Regular measurement and trending of impedance is therefore a fundamental part of every device check and remote monitoring alert configuration.
Right Answer: D
Quiz
Question 4/104/10
Leads and Lead Technology
Leads and Lead Technology
Leads and Lead Technology
What is the main advantage of an active-fixation lead compared with a passive-fixation tined lead?
Select the answer:Select the answer
1 correct answer
A.
It never causes perforation
B.
Placement is possible in sites lacking trabeculation, such as the septum, and extraction is generally easier
C.
It always has lower acute thresholds immediately after deployment
D.
It does not require a stylet
E.
It cannot dislodge under any circumstances
Explanation: Passive-fixation leads use tines or fins that become entangled in trabeculations, such as in the right ventricular apex or right atrial appendage, and are later secured by fibrosis. Active-fixation leads use an extendable-retractable or fixed helix that is screwed into the myocardium, allowing placement at almost any site, including the interventricular septum, right ventricular outflow tract, atrial septum, or in patients with smooth-walled or surgically altered chambers. Active-fixation leads are generally easier to extract after long dwell times because the tip has less extensive fibrotic encapsulation than tines, and the helix can sometimes be retracted. Disadvantages include a small risk of perforation, especially with thin-walled atria or when over-rotated, and a transient elevation in thresholds immediately after deployment owing to tissue injury, which usually improves within minutes. Both lead types can dislodge, although modern dislodgement rates are low. Most active-fixation leads require a stylet for positioning, while lumenless leads designed for conduction system pacing are delivered through a sheath. Lead choice depends on anatomy, intended site and operator preference.
Right Answer: B
Quiz
Question 5/105/10
Pacing Indications and Guidelines
Pacing Indications and Guidelines
Pacing Indications and Guidelines
An 82-year-old asymptomatic man has an incidental finding of Mobitz type II second-degree atrioventricular block with a broad QRS. Reversible causes are excluded. What is the most appropriate management?
Select the answer:Select the answer
1 correct answer
A.
No intervention unless symptoms develop
B.
Repeat ECG in 12 months
C.
Start a beta blocker
D.
Implantable loop recorder only
E.
Permanent pacing
Explanation: Mobitz type II second-degree block, advanced second-degree block and third-degree block not caused by reversible factors are indications for permanent pacing regardless of symptoms in current European guidelines. These conduction disturbances indicate infranodal disease and carry a risk of sudden progression to complete heart block with slow, unreliable escape rhythms, leading to syncope or sudden death. A broad QRS further supports infranodal involvement. Reversible causes, including drugs such as beta blockers, calcium channel blockers, digoxin or antiarrhythmic agents, electrolyte disturbances, acute myocardial infarction, Lyme disease and hypothyroidism, should first be excluded or corrected. Waiting for symptoms risks a first presentation with syncope and injury. A loop recorder is useful when symptoms are intermittent and the diagnosis uncertain, but not when a clear indication is already documented. Starting a beta blocker would worsen conduction. The choice of device, such as dual-chamber pacing for patients in sinus rhythm or consideration of conduction system pacing, depends on comorbidities and the expected ventricular pacing burden.
Right Answer: E
Quiz
Question 6/106/10
Pacing Modes and Timing Cycles
Pacing Modes and Timing Cycles
Pacing Modes and Timing Cycles
In the NBG pacemaker code, what does the third letter indicate?
Select the answer:Select the answer
1 correct answer
A.
The response to a sensed event
B.
The chamber paced
C.
The chamber sensed
D.
Rate modulation
E.
Multisite pacing
Explanation: The NASPE/BPEG generic (NBG) pacemaker code describes pacing modes using up to five positions. The first letter indicates the chamber or chambers paced: A for atrium, V for ventricle, D for dual, O for none. The second letter indicates the chamber or chambers sensed using the same letters. The third letter indicates the response to a sensed event: I for inhibited, where a sensed event inhibits pacing output; T for triggered, where a sensed event triggers a pacing output; D for dual, meaning both inhibition and triggering, such as atrial sensing triggering ventricular pacing after the atrioventricular delay while ventricular sensing inhibits; and O for no response, as in asynchronous modes. The fourth position indicates rate modulation, with R signifying a sensor-driven rate. The fifth position indicates multisite pacing in the atrium, ventricle or both. For example, DDDR means pacing and sensing in both chambers with dual response and rate modulation. Understanding this code is essential for interpreting programmed modes and describing device behaviour accurately in clinical communication.
Right Answer: A
Quiz
Question 7/107/10
Pacemaker Algorithms and Special Features
Pacemaker Algorithms and Special Features
Pacemaker Algorithms and Special Features
A patient with a dual-chamber pacemaker has a regular wide-complex paced tachycardia at the upper tracking rate. The marker channel shows retrograde atrial sensed events following each ventricular paced beat at a constant interval. What is the diagnosis?
Select the answer:Select the answer
1 correct answer
A.
Sinus tachycardia tracking
B.
Atrial fibrillation with mode switching
C.
Ventricular tachycardia
D.
Sensor-driven tachycardia
E.
Pacemaker-mediated (endless-loop) tachycardia
Explanation: Pacemaker-mediated tachycardia, also known as endless-loop tachycardia, occurs in dual-chamber tracking modes when a ventricular paced beat conducts retrogradely to the atrium, producing a P wave that is sensed outside the post-ventricular atrial refractory period. The device then tracks this retrograde P wave and paces the ventricle after the atrioventricular delay, and the cycle repeats. The rate is typically at or near the upper tracking rate because the sum of the atrioventricular delay and retrograde conduction time is usually shorter than the upper rate interval. Triggers include premature ventricular complexes, loss of atrial capture, atrial undersensing and magnet removal. The marker channel shows atrial sensed events at a constant ventriculoatrial interval after each ventricular pace. Most devices include algorithms that detect a sustained tachycardia at the upper rate with stable ventriculoatrial intervals and then extend the post-ventricular atrial refractory period for one beat or withhold a ventricular pace to break the loop. Magnet application terminates it acutely by disabling sensing. Prevention involves an adequate post-ventricular atrial refractory period.
Right Answer: E
Quiz
Question 8/108/10
Pacemaker Follow-up and Troubleshooting
Pacemaker Follow-up and Troubleshooting
Pacemaker Follow-up and Troubleshooting
A rhythm strip from a VVI pacemaker shows pacing spikes that are not followed by QRS complexes, while the timing of spikes is appropriate. What is this called?
Select the answer:Select the answer
1 correct answer
A.
Undersensing
B.
Oversensing
C.
Crosstalk
D.
Loss of capture
E.
Pacemaker-mediated tachycardia
Explanation: Loss of capture occurs when a pacing stimulus is delivered but fails to depolarise the myocardium. On the ECG, a pacing spike appears without a subsequent paced complex, either intermittently or persistently, while the timing of spikes remains appropriate, indicating normal sensing and timing function. Causes include a rise in capture threshold due to lead dislodgement or micro-dislodgement, exit block caused by fibrosis or inflammation at the electrode-tissue interface, perforation, conductor fracture or insulation breach, electrolyte disturbances such as hyperkalaemia, drugs that raise thresholds such as flecainide, myocardial infarction at the lead site, battery depletion and inappropriate programming with insufficient output. Functional non-capture occurs when a stimulus falls in the refractory period of a recently depolarised myocardium and should not be confused with true loss of capture. Management involves interrogating the device, measuring thresholds and impedance, reviewing chest X-ray for lead position, correcting metabolic abnormalities and increasing output or revising the lead as needed. Pacemaker-dependent patients with loss of capture require urgent attention.
Right Answer: D
Quiz
Question 9/109/10
ICD Therapy and Programming
ICD Therapy and Programming
ICD Therapy and Programming
Which patient meets a widely accepted primary prevention indication for an implantable cardioverter defibrillator?
Select the answer:Select the answer
1 correct answer
A.
Ejection fraction 45 percent with no symptoms
B.
Ejection fraction 30 percent three days after myocardial infarction
C.
NYHA class IV heart failure, not a transplant or device candidate, with life expectancy of three months
D.
Ischaemic cardiomyopathy with ejection fraction 30 percent, NYHA class II, more than 40 days after myocardial infarction on optimal medical therapy
E.
Vasovagal syncope with normal heart
Explanation: Primary prevention defibrillators aim to prevent sudden death in patients who have not yet had a sustained ventricular arrhythmia but are at high risk. Trials such as MADIT II and SCD-HeFT established benefit in patients with left ventricular ejection fraction of 35 percent or less and NYHA class II to III symptoms despite at least three months of optimal medical therapy, particularly in ischaemic cardiomyopathy. Implantation is deferred until at least 40 days after myocardial infarction and 90 days after revascularisation, because trials such as DINAMIT and IRIS showed no mortality benefit in the early post-infarction period, when ejection fraction may recover. Patients with non-ischaemic cardiomyopathy derive less clear benefit, as shown by the DANISH trial, and decisions involve age and comorbidities. Defibrillators are not recommended for patients with NYHA class IV symptoms refractory to treatment unless they are candidates for resynchronisation, ventricular assist devices or transplantation, nor for those with life expectancy under one year. Ejection fraction above 35 percent without other risk factors, and vasovagal syncope, are not indications.
Right Answer: D
Quiz
Question 10/1010/10
ICD Electrogram Interpretation
ICD Electrogram Interpretation
ICD Electrogram Interpretation
A stored defibrillator episode shows a tachycardia with ventricular rate 190 per minute, atrial rate 95 per minute and atrioventricular dissociation, with a change in near-field morphology compared with sinus rhythm. What is the rhythm?
Select the answer:Select the answer
1 correct answer
A.
Sinus tachycardia
B.
Ventricular tachycardia
C.
Atrial flutter with 2:1 conduction
D.
Atrioventricular nodal re-entrant tachycardia
E.
Atrial fibrillation
Explanation: When analysing defibrillator electrograms, the relationship between atrial and ventricular activity is a key discriminating feature. A ventricular rate greater than the atrial rate with atrioventricular dissociation is diagnostic of ventricular tachycardia, because supraventricular tachycardias require atrial or junctional activation conducting to the ventricles, giving atrial rates equal to or greater than ventricular rates. Here the ventricular rate of 190 per minute is double the atrial rate of 95 per minute, with no fixed relationship, indicating the ventricle is driven independently. The change in morphology of the ventricular electrogram compared with the sinus template supports a ventricular origin. Sinus tachycardia and nodal re-entry would show 1:1 relationships with stable atrioventricular timing. Atrial flutter with 2:1 conduction would show an atrial rate twice the ventricular rate, not the reverse. Atrial fibrillation would show rapid irregular atrial activity with irregular ventricular response. Recognising this pattern confirms appropriate therapy and guides further management, such as antiarrhythmic drugs or ablation for recurrent ventricular tachycardia.
BHRS202110DE: BHRS_Devices Practice test unlocks all online simulator questions
Thank you for choosing the free version of the BHRS202110DE: BHRS_Devices practice test! Further deepen your knowledge on British Heart Rhythm Society (BHRS) Simulator; by unlocking the full version of our BHRS202110DE: BHRS_Devices Simulator you will be able to take tests with over 108 constantly updated questions and easily pass your exam. 98% of people pass the exam in the first attempt after preparing with our 108 questions.
What to expect from our BHRS202110DE: BHRS_Devices practice tests and how to prepare for any exam?
The BHRS202110DE: BHRS_Devices Simulator Practice Tests are part of the British Heart Rhythm Society (BHRS) Database and are the best way to prepare for any BHRS202110DE: BHRS_Devices exam. The BHRS202110DE: BHRS_Devices practice tests consist of 108 questions and are written by experts to help you and prepare you to pass the exam on the first attempt. The BHRS202110DE: BHRS_Devices database includes questions from previous and other exams, which means you will be able to practice simulating past and future questions. Preparation with BHRS202110DE: BHRS_Devices Simulator will also give you an idea of the time it will take to complete each section of the BHRS202110DE: BHRS_Devices practice test . It is important to note that the BHRS202110DE: BHRS_Devices Simulator does not replace the classic BHRS202110DE: BHRS_Devices study guides; however, the Simulator provides valuable insights into what to expect and how much work needs to be done to prepare for the BHRS202110DE: BHRS_Devices exam.
BHRS202110DE: BHRS_Devices Practice test therefore represents an excellent tool to prepare for the actual exam together with our British Heart Rhythm Society (BHRS) practice test . Our BHRS202110DE: BHRS_Devices Simulator will help you assess your level of preparation and understand your strengths and weaknesses. Below you can read all the quizzes you will find in our BHRS202110DE: BHRS_Devices Simulator and how our unique BHRS202110DE: BHRS_Devices Database made up of real questions:
Info quiz:
Quiz name:BHRS202110DE: BHRS_Devices
Total number of questions:108
Number of questions for the test:50
Pass score:80%
You can prepare for the BHRS202110DE: BHRS_Devices exams with our mobile app. It is very easy to use and even works offline in case of network failure, with all the functions you need to study and practice with our BHRS202110DE: BHRS_Devices Simulator.
Use our Mobile App, available for both Android and iOS devices, with our BHRS202110DE: BHRS_Devices Simulator . You can use it anywhere and always remember that our mobile app is free and available on all stores.
Our Mobile App contains all BHRS202110DE: BHRS_Devices practice tests which consist of 108 questions and also provide study material to pass the final BHRS202110DE: BHRS_Devices exam with guaranteed success.
Our BHRS202110DE: BHRS_Devices database contain hundreds of questions and British Heart Rhythm Society (BHRS) Tests related to BHRS202110DE: BHRS_Devices Exam. This way you can practice anywhere you want, even offline without the internet.